Abstract
STUDY DESIGN: Single case report and extensive literature review. OBJECTIVES: To present the first such report of cervical cord contusion after the percutaneous placement of gold-seed fiducials. The pathomechanics and surgical recommendations are reviewed. BACKGROUND: Spinal cord injuries are well documented in the medical literature. These injuries range from cord contusion to transection and result primarily from trauma. A single case report of a patient who was found to have a nonhemorrhagic cervical spinal cord contusion after percutaneous fiducial implantation is presented. METHODS: Single case report. RESULTS: The patient underwent percutaneous placement of fiducials for stereotactic radiosurgery for a nerve sheath tumor. Postoperatively she had primarily sensory complaints; no motor deficits were detected on neurological examination. Neuroimaging studies demonstrated nonhemorrhagic cervical cord contusion. She was treated conservatively and had complete resolution of her symptoms. CONCLUSIONS: The likely mechanism for the contusion was neck hyperextension during thrusting maneuvers during fiducial implantation. This is yet another report of normal intraoperative-evoked potentials with postoperative neurological sequelae. A dedicated team approach involving ancillary staff, anesthesiologists, and surgeons should be utilized to avert this potentially devastating complication.