Accuracy of shoulder ultrasound examination for diagnosis of rotator cuff pathologies: a single-center retrospective study

肩关节超声检查诊断肩袖病变的准确性:一项单中心回顾性研究

阅读:1

Abstract

BACKGROUND: Shoulder pathologies need accurate diagnosis for best management and treatment provided to patients. OBJECTIVE: Determine the diagnostic sensitivity, specificity and accuracy of shoulder ultrasonography (US). DESIGN: Retrospective, analytical. SETTING: Tertiary care center in Riyadh, Saudi Arabia. PATIENTS AND METHODS: We included all shoulder exams performed between January 2010 and December 2016 that met the inclusion criteria. Data was collected retrospectively from the a picture archiving and communication system and patient records. The patients were evaluated using US for the presence of rotator cuff tears and classified into intact, full-thickness tear, partial-thickness tear, tendinosis, subacromial/subdeltoid bursitis and acromioclavicular joint degenerative changes. The US findings were correlated with the shoulder MRI study findings. The time interval between the US examination and MRI ranged from 0 to 180 days (6 months). MAIN OUTCOME MEASURES: To compare the sensitivity, specificity and accuracy of shoulder US studies in the detection of rotator cuff pathologies in comparison to MRI findings. SAMPLE SIZE AND CHARACTERISTICS: 86 (60 females, 26 male), mean age 53.7 years (range, 19-85). RESULTS: The sensitivity, specificity and accuracy of US for the detection of full-thickness supraspinatus tears compared with those of MRI were 86%, 82% and 83%, respectively. The sensitivity, specificity and accuracy of US for the detection of partial-thickness supraspinatus tears compared with those of MRI were 38%, 70% and 58%, respectively. Overall PPV, NPV, sensitivity, specificity and accuracy of US for the detection of full-thickness tears compared with those of MRI were 35%, 97%, 78%, 83% and 83%, respectively. For partial-thickness tears, the overall PPV, NPV, sensitivity, specificity and accuracy of US compared with those of MRI were 51%, 60%, 51%, 60% and 56%, respectively. CONCLUSION: Overall, US has high sensitivity, specificity and accuracy for the detection of full-thickness tears compared with the detection of partial-thickness tears. LIMITATIONS: Small sample size and retrospective. CONFLICT OF INTEREST: None.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。