Development and validation of a new HPLC analytical method for the determination of diclofenac in tablets

开发和验证一种测定片剂中双氯芬酸含量的新型高效液相色谱分析方法

阅读:1

Abstract

A new selective and sensitive high-performance liquid chromatography (HPLC) method was developed for the quantification of diclofenac sodium (DS) in pharmaceutical dosage form using lidocaine as internal standard (IS). Chromatographic separation was achieved on a symmetry C18 column (4.6 mm × 150 mm, 3 μm spherical particles) using 0.05 M orthophosporic (pH 2.0) 35% and acetonitrile as 65%, as the mobile phase at a flow rate of 2.0 mL/min and monitored at 210 nm. The run time was 2 min. The method was validated to fulfill International Conference on Harmonisation (ICH) requirements and this validation included specificity, linearity, limit of detection (LOD), limit of quantification (LOQ), accuracy, precision and robustness. The calibration curve was linear over the concentration range from 10 to 200 µg/ml, and lower limit of detection of 12.5 ng/ml. The accuracy and precision of the method were within the acceptable limit of ±20% at the lower limit of quantitation and ±15% at other concentrations. Diclofenac was unstable at room temperature it showed more than 25% loss after 24 h. While, DS is very stable at refrigerator 4 °C auto-sampler, freeze/thaw cycles and 30 days storage in a freezer at -35 ± 2 °C. All results were acceptable and this confirmed that the method is suitable for its intended use in routine quality control and assay of drugs.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。