Transluminal removal of a fractured and embolized indwelling central venous catheter in the pulmonary artery

经腔内取出肺动脉内断裂栓塞的留置中心静脉导管

阅读:1

Abstract

Vascular catheters are associated with complications such as infection, thrombosis and stenosis. The embolization of a venous catheter fragment is a rare complication. A 39-year-old woman underwent placement of a totally implantable venous access device for chemotherapy to treat a recurrent liposarcoma of the left thigh. The "pinch-off sign" was noted on a routine chest X-ray but that was ignored. Three-months after implantation of the intravenous access device, the indwelling central catheter was fractured and embolized to the pulmonary trunk. The catheter in the pulmonary trunk was successfully removed through a percutaneous femoral vein approach using a pigtail catheter and goose neck snare.