Two dislodged and crushed coronary stents: treatment of two simultaneously dislodged stents using crushing techniques

两枚脱落并压碎的冠状动脉支架:采用压碎技术治疗两枚同时脱落的支架

阅读:1

Abstract

Coronary stent dislodgement is a rare complication of percutaneous coronary intervention. We report a rare case of dislodgement of two intracoronary stents. On withdrawal of two balloon catheters, one with a guide wire was mechanically distorted from the left main (LM) to the proximal left anterior descending artery (LAD) while the other was dislodged from the LM to the ostial left circumflex artery. The stent in the LAD could not be retrieved into the guide catheter using a Goose neck snare, because it was caught on a previously deployed stent at the mid LAD. A new stent was quickly deployed from the LM to the proximal LAD, because the patient developed cardiogenic shock. Both stents, including a distorted and elongated stent, were crushed to the LM wall. Stent deployment and crushing may be a good alternative technique to retrieving a dislodged stent.