A Prospective Clinical Study to EvaluAte the AbiliTy of the CloudCath System to Detect Peritonitis During In-Home Peritoneal Dialysis (CATCH)

一项前瞻性临床研究,旨在评估 CloudCath 系统在家庭腹膜透析期间检测腹膜炎的能力 (CATCH)

阅读:1

Abstract

INTRODUCTION: Peritonitis is the leading complication of peritoneal dialysis (PD). Patients are instructed to seek care promptly for signs (cloudy effluent) or symptoms (abdominal pain), and earlier treatment improves outcomes. The CloudCath Peritoneal Dialysis Drain Set Monitoring (CloudCath) system monitors turbidity in dialysis effluent and sends notifications of changes signaling possible peritonitis. METHODS: We conducted this single-arm, open-label, multicenter study of CloudCath system use during PD. We deactivated system notifications to participants and investigators, who followed standard-of-care for peritonitis signs and symptoms. Effectiveness endpoints measured time between CloudCath system notifications and peritonitis events using International Society of Peritoneal Dialysis (ISPD) criteria. RESULTS: Two hundred forty-three participants used the CloudCath system for 178.8 patient-years. Of 71 potential peritonitis events, 51 events (0.29 per patient-year) met ISPD white blood cell (WBC) count criteria. The system triggered notifications for 41 of 51 events (80.4%), with a median lead time of 2.6 days (10%-90% range, -1.0 to 15.7; P < 0.0001). Excluding 6 peritonitis events that occurred when the system was not in use, the system triggered notifications for 41 of 45 events (91.1%), with a median lead time of 3.0 days (10%-90% range, -0.5 to 18.8; P < 0.0001). Of the 0.78 notifications per patient-year, the majority were peritonitis events or nonperitonitis events such as exit site and tunnel infections or catheter/cycler issues. CONCLUSION: The CloudCath system detected peritonitis events during PD several days earlier than the current standard-of-care and has the capacity to send notifications that could expedite peritonitis diagnosis and treatment.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。