Improving decomposition image quality in dual-energy chest radiography using two-dimensional crisscrossed anti-scatter grid

利用二维交叉防散射栅格提高双能胸部X线摄影的分解图像质量

阅读:1

Abstract

BACKGROUND: Chest radiography is a widely used medical imaging modality for diagnosing chest-related diseases. However, anatomical structure overlap hinders accurate lesion detection. While the dual-energy x-ray imaging technique addresses this issue by separating soft-tissue and bone images from an original chest radiograph, scattered radiation remains a significant challenge in decomposition image quality. PURPOSE: This work aims to conduct dual-energy material decomposition (DEMD) in chest radiography using a two-dimensional (2D) crisscrossed anti-scatter grid to improve decomposition image quality by effectively removing scattered radiation. METHODS: A 2D graphite-interspaced grid with a strip density of N = 1.724 lines/mm and grid ratio r = 6:1 was fabricated using a high-precision sawing process. The grid characteristics were evaluated using the IEC standard fixture. A 2D-grid-based DEMD process, which involves the acquisition of low- and high-kV radiographs with a 2D grid, generation of a pairwise decomposition function using a calibration wedge phantom, and decomposition of soft-tissue and bone images using the decomposition function, was implemented, followed by software-based grid artifact reduction. Experiments were conducted on a commercially available chest phantom using an x-ray imaging system operating at two tube voltages of 70 and 120 kVp. The decomposition image quality of the proposed DEMD and conventional dual-energy subtraction methods was compared for the cases of no grid, software-based scatter correction, 1D grid (N = 8.475 lines/mm and r = 12:1), and 2D grid. RESULTS: The 2D grid demonstrated superior scatter radiation removal ability with scatter radiation transmission of 6.34% and grid selectivity of 9.67, representing a 2.6-fold decrease and a 2.7-fold improvement over the 1D grid, respectively. Compared to other competitive methods, the 2D-grid-based DEMD method considerably improved decomposition image quality, with improved lung structure visibility in selective soft-tissue images. CONCLUSIONS: The proposed DEMD method yielded high-quality dual-energy chest radiographs by effectively removing scattered radiation, demonstrating significant potential for improving lesion detection in clinical practice.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。