Comparison of central corneal thickness using anterior segment optical coherence tomography vs ultrasound pachymetry

前节光学相干断层扫描与超声角膜测厚法测量中央角膜厚度的比较

阅读:1

Abstract

PURPOSE: To determine if there is a systematic difference in central corneal thickness (CCT) measured using anterior segment optical coherence tomography (AS-OCT) as compared with ultrasound pachymetry. DESIGN: Prospective observational study. METHODS: Consecutive subjects with clinically normal corneas underwent CCT measurement by both ultrasound and AS-OCT while participating in a population-based study in Ghana, West Africa. One eye of each subject was randomly selected for analysis. Two measurements were taken and averaged. Agreement and interobserver variability were also analyzed. RESULTS: One hundred and fifty-five subjects of African ethnicity and average age 57 years (standard deviation [SD] 12; range, 40 to 98 years) were included. Measurements by AS-OCT and US were taken a mean of 15 days (maximum, six weeks) apart. The mean (SD) [range] US CCT was 525.3 microm (33.5) [422, 653] and 499.0 microm (32.0) [428, 613] with AS-OCT. Measurements by the two modalities were strongly correlated (r(2) = 0.82; P < .001), and a significant difference was observed between mean US and AS-OCT CCT (SD) [range] of 26.3 microm (14.2) [-63, 12] (P < .001). The width of the limits of agreement was 28 microm, about 6% of the average pachymetry reading. In 50 eyes randomly remeasured with OCT by a second observer, the intraclass correlation coefficient was 0.91. There was a small but significant systematic difference between observers (mean 6.9 microm, SD 10.9 microm), or 1.4% (P < .001), increasing the difference noted above. CONCLUSION: There is a reproducible systematic difference between CCT measurements taken with ultrasound and OCT. It is important to note in clinical practice, that measurements acquired by these two modalities are not directly interchangeable.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。