Radiation Dose Reduction in Dual-Energy CT: Does It Affect the Accuracy of Urinary Stone Characterization?

双能CT辐射剂量降低:是否会影响泌尿系结石鉴别的准确性?

阅读:2

Abstract

OBJECTIVE: The purpose of this article is to assess the effect of radiation dose reduction in dual-energy CT (DECT) on the performance of renal stone characterization using a patient cohort. MATERIALS AND METHODS: CT data from 39 unenhanced DECT examinations performed for stone characterization were retrospectively analyzed in this study. Reduced-dose images were simulated at 75%, 50%, and 25% of the routine dose using a previously validated noise-insertion algorithm. Differentiation between uric acid (UA) and non-UA stones was performed using a fixed cutoff value for the dual-energy ratio. ROC analysis was performed to determine optimal cutoff values and the associated sensitivity and specificity. RESULTS: Of the 206 stones found, 43 were UA and 163 were non-UA. The mean (± SD) volume CT dose index (CTDIvol) was 16.0 ± 4.0 mGy at the 100% dose level. The mean noise in 100-kV images increased from 40.9 ± 6.8 HU at 100% dose to 46.8 ± 8.8 HU, 57.7 ± 12.5 HU, and 85.4 ± 22.9 HU at 75%, 50%, and 25% dose levels, respectively. Using the default cutoff value, for stones 10 mm(3) or larger, the sensitivity/specificity were 100.0%/98.8%, 82.8%/98.8%, and 89.3%/98.7%, at 75%, 50%, and 25% dose levels, respectively. ROC analysis showed varying optimal cutoff values at different dose levels. The sensitivity and specificity improved with use of these optimal cutoff values. Differentiation capability decreased for stones smaller than 10 mm(3). CONCLUSION: At 75% of the 16-mGy routine dose, the sensitivity and specificity for differentiating UA from non-UA stones were minimally affected for stones 10 mm(3) or larger. The use of optimal cutoff values for dual-energy ratio as dose decreased (and noise increased) provided improved performance.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。