Quantification of hepatic blood flow using a high-resolution phase-contrast MRI sequence with compressed sensing acceleration

利用高分辨率相位对比磁共振成像序列和压缩感知加速技术对肝脏血流量进行定量分析

阅读:1

Abstract

OBJECTIVE. The objective of our study was to evaluate the performance of a high-spatial-resolution 2D phase-contrast (PC) MRI technique accelerated with compressed sensing for portal vein (PV) and hepatic artery (HA) flow quantification in comparison with a standard PC MRI sequence. SUBJECTS AND METHODS. In this prospective study, two PC MRI sequences were compared, one with parallel imaging acceleration and low spatial resolution (generalized autocalibrating partial parallel acquisition [GRAPPA]) and one with compressed sensing acceleration and high spatial resolution (sparse). Seventy-six patients were assessed, including 37 patients with cirrhosis. Two observers evaluated PC image quality. Quantitative analyses yielded a mean velocity, flow, and vessel area for the PV and HA and an arterial fraction. The PC techniques were compared using the paired Wilcoxon test and Bland-Altman statistics. The sensitivity of the flow parameters to the severity of cirrhosis was also assessed. RESULTS. Vessel delineation was significantly improved using the PC sparse sequence (p < 0.034). For both in vitro and in vivo measurements, PC sparse yielded lower estimates for vessel area and flow, and larger differences between PC GRAPPA and PC sparse were observed in the HA. PV velocity and flow were significantly lower in patients with cirrhosis on both PC sparse (p < 0.001 and p = 0.042, respectively) and PC GRAPPA (p < 0.001 and p = 0.005, respectively). PV velocity correlated negatively with Child-Pugh class (r = -0.50, p < 0.001), whereas the arterial fraction measured with PC sparse was higher in patients with Child-Pugh class B or C disease than in those with Child-Pugh class A disease, with a trend toward significance (p = 0.055). CONCLUSION. A high-spatial-resolution highly accelerated compressed sensing technique (PC sparse) allows total hepatic blood flow measurements obtained in 1 breath-hold, provides improved delineation of the hepatic vessels compared with a standard PC MRI sequence (GRAPPA), and can potentially be used for the noninvasive assessment of liver cirrhosis.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。