Comparison of radiation dose and image quality: 320-MDCT versus 64-MDCT coronary angiography

辐射剂量和图像质量比较:320层螺旋CT与64层螺旋CT冠状动脉造影

阅读:1

Abstract

OBJECTIVE: The purpose of this study was to compare radiation dose and image quality of 320- and 64-MDCT angiography using prospective gating. MATERIALS AND METHODS: One hundred seventy-four patients underwent 320-MDCT, and 95 patients underwent 64-MDCT. The scan parameters for 320-MDCT were 120 kVp, 400 mA, and gantry rotation of 350 milliseconds; the parameters for 64-MDCT were 120 kVp, 600 mA, and gantry rotation of 350 milliseconds. Effective dose (ED) was calculated from the dose-length product and a conversion factor (k = 0.014 mSv / mGy × cm). Two observers independently assessed image quality using a 3-point scale, where 3 denotes excellent quality and 1 denotes nondiagnostic quality, using a 16-segment model. Discrepancies were settled by consensus. RESULTS: The ED was significantly lower in patients undergoing 320-MDCT angiography, with a median ED of 4.4 mSv (interquartile range [IQR], 3.4-6.2 mSv), compared with 64-MDCT angiography, with a median ED of 6.2 mSv (IQR, 5.5-6.9 mSv) (p = 0.0001). In patients with a heart rate of 65 beats/min or less (92%), the median radiation dose using 320-MDCT was 4.1 mSv (IQR, 3.2-6.1 mSv), and that for 64-MDCT angiography was 6.2 mSv (IQR, 5.8-6.9 mSv) (p = 0.0001). In patients with heart rate greater than 65 beats/min (8%), the median dose was higher with 320-MDCT (8.7 mSv; IQR, 5.9-14.3 mSv) than with 64-MDCT (5.8 mSv; IQR, 5.3-6.7 mSv) (p = 0.02). Segmental image quality was significantly better for 320-MDCT (excellent or good quality, 96.66%; nondiagnostic quality, 0.1%) than for 64-MDCT angiography (excellent or good quality, 86%; nondiagnostic quality, 3.33%) (all p < 0.0001). CONCLUSION: Image quality was good for both 320- and 64-MDCT angiography. Overall radiation dose was significantly lower in 320-MDCT angiography when the heart rate was 65 beats/min or less. Every effort should be made to control heart rate to minimize radiation dose.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。