High frequency label-free photoacoustic microscopy of single cells

高频无标记单细胞光声显微镜

阅读:1

Abstract

Photoacoustic measurements of melanoma cells and red blood cells (RBCs) using ultra-high frequency (UHF) wide-bandwidth transducers are reported. In this detection system, the resolution typically depends on the parameters of the receiving transducer, and not the focus of the laser. A single melanoma cell was imaged with 200, 375 and 1200 MHz transducers. As the frequency increased, the resolution increased, resulting in greater detail observed. A single RBC was imaged at 1200 MHz, showing the contours of the cell. While lateral and axial resolutions approaching 1 μm are possible with this microscope, the key advantage is the ability to perform a wide-bandwidth quantitative signal analysis of the photoacoustic signals. The power spectrum of the signals measured from RBCs showed distinct spectral minima around 800 and 1500 MHz which are directly related to the RBC geometry. This study reports on the high-resolution imaging capabilities and quantitative analyses using UHF photoacoustic microscopy.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。