Ppb-level gas detection using on-beam quartz-enhanced photoacoustic spectroscopy based on a 28 kHz tuning fork

利用基于28 kHz音叉的光束石英增强光声光谱技术实现ppb级气体检测

阅读:1

Abstract

In this paper, an on-beam quartz-enhanced photoacoustic spectroscopy (QEPAS) sensor based on a custom quartz tuning fork (QTF) acting as a photoacoustic transducer, was realized and tested. The QTF is characterized by a resonance frequency of 28 kHz, ~15% lower than that of a commercially available 32.7 kHz standard QTF. One-dimensional acoustic micro resonator (AmR) was designed and optimized by using stainless-steel capillaries. The 28 kHz QTF and AmRs are assembled in on-beam QEPAS configuration. The AmR geometrical parameters have been optimized in terms of length and internal diameter. The laser beam focus position and the AmR coupling distance were also adjusted to maximize the coupling efficiency. For comparison, QEPAS on-beam configurations based on a standard QTF and on the 28 kHz QTF were compared in terms of H(2)O and CO(2) detection sensitivity. In order to better characterize the performance of the system, H(2)O, C(2)H(2) and CO(2) were detected for a long time and the long-term stability was analyzed by an Allan variance analysis. With the integration time of 1 s, the detection limits for H(2)O, C(2)H(2) and CO(2) are 1.2 ppm, 28.8 ppb and 2.4 ppm, respectively. The detection limits for H(2)O, C(2)H(2) and CO(2) can be further improved to 325 ppb, 10.3 ppb and 318 ppb by increasing the integration time to 521 s, 183 s and 116 s.