Repeatability of short-duration transient visual evoked potentials in normal subjects

正常受试者短时程瞬态视觉诱发电位的重复性

阅读:1

Abstract

To evaluate the within-session and inter-session repeatability of a new, short-duration transient visual evoked potential (SD-tVEP) device on normal individuals, we tested 30 normal subjects (20/20 visual acuity, normal 24-2 SITA Standard VF) with SD-tVEP. Ten of these subjects had their tests repeated within 1-2 months from the initial visit. Synchronized single-channel EEG was recorded using a modified Diopsys Enfant System (Diopsys, Inc., Pine Brook, New Jersey, USA). A checkerboard stimulus was modulated at two reversals per second. Two different contrasts of checkerboard reversal patterns were used: 85% Michelson contrast with a mean luminance of 66.25 cd/m(2) and 10% Michelson contrast with a mean luminance of 112 cd/m(2). Each test lasted 20 s. Both eyes, independently and together, were tested 10 times (5 times at each contrast level). The following information was identified from the filtered N75-P100-N135 complex: N75 amplitude, N75 latency, P100 amplitude, P100 latency, and Delta Amplitude (N75-P100). The median values for each eye's five SD-tVEP parameters were calculated and grouped into two data sets based on contrast level. Mean age was 27.3 +/- 5.2 years. For OD only, the median (95% confidence intervals) of Delta Amplitude (N75-P100) amplitudes at 10% and 85% contrast were 4.6 uV (4.1-5.9) and 7.1 uV (5.15-9.31). The median P100 latencies were 115.2 ms (112.0-117.7) and 104.0 ms (99.9-106.0). There was little within-session variability for any of these parameters. Intraclass correlation coefficients ranged between 0.64 and 0.98, and within subject coefficients of variation were 3-5% (P100 latency) and 15-30% (Delta Amplitude (N75-P100) amplitude). Bland-Altman plots showed good agreement between the first and fifth test sessions (85% contrast Delta Amplitude (N75-P100) delta amplitude, mean difference, 0.48 mV, 95% CI, -0.18-1.12; 85% contrast P100 latency delay, -0.82 ms, 95% CI, -3.12-1.46; 10% contrast Delta Amplitude (N75-P100) amplitude, 0.58 mV, 95% CI, -0.27-1.45; 10% contrast P100 latency delay, -2.05 mV, 95% CI, -5.12-1.01). The inter-eye correlation and agreement were significant for both SD-tVEP amplitude and P100 latency measurements. For the subset of eyes in which the inter-session repeatability was tested, the intraclass correlation coefficients ranged between 0.71 and 0.86 with good agreement shown on Bland-Altman plots. Short-duration transient VEP technology showed good within-session, inter-session repeatability, and good inter-eye correlation and agreement.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。