The Humphrey optical coherence tomography scanner: quantitative analysis and reproducibility study of the normal human retinal nerve fibre layer

Humphrey光学相干断层扫描仪:正常人视网膜神经纤维层的定量分析和重复性研究

阅读:1

Abstract

BACKGROUND/AIMS: To determine the reproducibility of the Humphrey optical coherence tomography scanner (OCT), software version 5.0, for measurement of retinal nerve fibre layer (RNFL) thickness in normal subjects and to compare OCT measurements with published histological thickness of the human RNFL. METHODS: Three independent measurements were obtained at each session for one eye from 15 normal subjects with a mean age of 30.8 (SD 10.9) years. Scans were taken in the peripapillary retina using the default setting (1.74 mm radius from centre of the optic disc) and were repeated 1 week later. Additional scans were obtained at the optic nerve head (ONH) margin overlying the scleral rim, for comparison with available histological data on the human RNFL. RESULTS: For the 1.74 mm circular scan, the mean coefficient of variation (COV) for the global RNFL thickness measurement was 5% (SD 3%). This increased to 8% (3%) for quadrant measurements and to 9% (3%) with further subdivision into 12 segments. Significant differences (p<0.05) between sessions were only found when the data were divided into segments. The mean RNFL thickness for the 1.74 mm scan was 127.87 (9.81) microm. The RNFL was maximal at the superior disc pole, 161.44 microm (14.8), and minimal at the temporal pole, 83.1 (12.8) microm. Peak thickness values occurred superior temporal and inferior temporal to the vertical axis. RNFL thickness for every sector of the disc was greatest at the margin of the optic disc (mean 185.79 microm; SD 32.61). Although the variation in RNFL thickness around the disc follows published histology data, the OCT underestimates RNFL thickness by an average of 37% (SD 11; range 21-48%). CONCLUSION: The OCT provides reproducible measurement of the retinal structures that are consistent with the properties of the RNFL. However, comparison with available studies of RNFL thickness in the human suggests that in its present form, the OCT underestimates RNFL thickness. Further refinement of this technology is required to improve the accuracy with which the OCT measures retinal nerve fibre layer thickness.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。