Accuracy of currently available neonatal respiratory function monitors for neonatal resuscitation

目前可用的新生儿呼吸功能监护仪在新生儿复苏中的准确性

阅读:1

Abstract

This study aimed to test the accuracy in volume measurements of three available respiratory function monitors (RFMs) for neonatal resuscitation and the effect of changing gas conditions. The Florian, New Life Box Neo-RSD (NLB Neo-RSD) and NICO RFM were tested on accuracy with volumes of 10 and 20 mL and on changes in volume measurements under changing gas conditions (oxygen level 21-100 % and from cold dry air (24 ± 2 °C) to heated humidified air (37 °C). Volume differences >10 % were considered clinically relevant. We found that the mean (SD) volume difference was clinically acceptable for all devices (10, 20 mL): Florian (+8.4 (1.2)%, +8.4 (0.5)%); NLB Neo-RSD (+5.8 (1.1)%, +4.3 (1.4)%); and NICO (-8.2 (0.9)%, -8.7 (0.8)%). Changing from cold dry to heated humidified air increased the volume difference using the Florian (cold dry air, heated humidified air (+5.2 (1.2)%, +12.2 (0.9)%) but not NLB Neo-RSD (+2.0(1.6)%, +3.4(2.8)%) and NICO (-2.3 % (0.8), +0.1 (0.6)%). Similarly, when using heated humidified air, increasing oxygen enlarged increased the volume difference using the Florian (oxygen 21 %, 100 %: +12.2(1.0)%, +19.8(1.1)%), but not NLB Neo-RSD (+0.2(1.9)%, +1.1(2.8)%) and NICO (-5.6(0.9)%, -3.7(0.9)%). Clinically relevant changes occurred when changing both gas conditions (Florian +25.7(1.7)%; NLB Neo-RSD +3.8(2.4)%; NICO -5.7(1.4)%). CONCLUSION: The available RFMs demonstrated clinically acceptable deviations in volume measurements, except for the Florian when changing gas conditions. WHAT IS KNOWN: •Respiratory function monitors (RFMs) are increasingly used for volume measurements during respiratory support of infants at birth. •During respiratory support at birth, gas conditions can change quickly, which can influence the volume measurements. What is new: •The available RFMs have clinically acceptable deviations when measuring the accuracy of volume measurements. •The RFM using a hot wire anemometer demonstrated clinically relevant deviations in volume measurements when changing the gas conditions. These deviations have to be taken into account when interpreting the volumes directly at birth.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。