Optofluidic laser for dual-mode sensitive biomolecular detection with a large dynamic range

用于大动态范围双模灵敏生物分子检测的光流控激光器

阅读:6
作者:Xiang Wu, Maung Kyaw Khaing Oo, Karthik Reddy, Qiushu Chen, Yuze Sun, Xudong Fan

Abstract

Enzyme-linked immunosorbent assay (ELISA) is a powerful method for biomolecular analysis. The traditional ELISA employing light intensity as the sensing signal often encounters large background arising from non-specific bindings, material autofluorescence and leakage of excitation light, which deteriorates its detection limit and dynamic range. Here we develop the optofluidic laser-based ELISA, where ELISA occurs inside a laser cavity. The laser onset time is used as the sensing signal, which is inversely proportional to the enzyme concentration and hence the analyte concentration inside the cavity. We first elucidate the principle of the optofluidic laser-based ELISA, and then characterize the optofluidic laser performance. Finally, we present the dual-mode detection of interleukin-6 using commercial ELISA kits, where the sensing signals are simultaneously obtained by the traditional and the optofluidic laser-based ELISA, showing a detection limit of 1 fg ml(-1) (38 aM) and a dynamic range of 6 orders of magnitude.