Beam hardening correction for computed tomography images using a postreconstruction method and equivalent tissue concept

利用后重建方法和等效组织概念对计算机断层扫描图像进行束硬化校正

阅读:1

Abstract

A postreconstruction method for correcting the beam-hardening artifacts in computed tomography (CT) images is proposed. This method does not require x-ray spectrum measurement. The authors assumed that a pixel in a CT image can be decomposed into equivalent tissue percentages, depending on its CT number. A scout view of the step wedges made of these equivalent tissues was performed to obtain a beam-hardening correction curve for each tissue. Projecting through the CT image from various angles generated simulated projection data and the total thickness of each tissue along the ray. The correction term was estimated using the tissue thickness traveled by the ray, and this term was then added to its corresponding projection data. A second reconstruction using the corrected projection data yielded a beam-hardening corrected image. The preliminary results show that this method reduces beam hardening artifacts by 14% for aluminum and increased the object contrast by 18% near the aluminum-water boundary. The variation in CT numbers at different locations were reduced, and the aluminum CT number also was restored.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。