Accuracy of an automatic patient-positioning system based on the correlation of two edge images in radiotherapy

基于放射治疗中两幅边缘图像相关性的自动患者定位系统的准确性

阅读:1

Abstract

We have clinically evaluated the accuracy of an automatic patient-positioning system based on the image correlation of two edge images in radiotherapy. Ninety-six head & neck images from eight patients undergoing proton therapy were compared with a digitally reconstructed radiograph (DRR) of planning CT. Two edge images, a reference image and a test image, were extracted by applying a Canny edge detector algorithm to a DRR and a 2D X-ray image, respectively, of each patient before positioning. In a simulation using a humanoid phantom, performed to verify the effectiveness of the proposed method, no registration errors were observed for given ranges of rotation, pitch, and translation in the x, y, and z directions. For real patients, however, there were discrepancies between the automatic positioning method and manual positioning by physicians or technicians. Using edged head coronal- and sagittal-view images, the average differences in registration between these two methods for the x, y, and z directions were 0.11 cm, 0.09 cm and 0.11 cm, respectively, whereas the maximum discrepancies were 0.34 cm, 0.38 cm, and 0.50 cm, respectively. For rotation and pitch, the average registration errors were 0.95° and 1.00°, respectively, and the maximum errors were 3.6° and 2.3°, respectively. The proposed automatic patient-positioning system based on edge image comparison was relatively accurate for head and neck patients. However, image deformation during treatment may render the automatic method less accurate, since the test image many differ significantly from the reference image.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。