Measurement of Signal-to-Noise and Contrast-to-Noise in the fBIRN Multicenter Imaging Study

fBIRN多中心影像研究中信噪比和对比噪声比的测量

阅读:1

Abstract

The ability to analyze and merge data across sites, vendors, and field strengths depends on one's ability to acquire images with the same image quality including image smoothness, signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR). SNR can be used to compare different magnetic resonance scanners as a measure of comparability between the systems. This study looks at the SNR and CNR ratios in structural fast spin-echo T2-weighted scans acquired in five individuals across ten sites that are part of Functional Imaging Research of Schizophrenia Testbed Biomedical Informatics Research Network (fBIRN). Different manufacturers, field strengths, gradient coils, and RF coils were used at these sites. The SNR of gray matter was fairly uniform (41.3-43.3) across scanners at 1.5 T. The higher field scanners produced images with significantly higher SNR values (44.5-108.7 at 3 T and 50.8 at 4 T). Similar results were obtained for CNR measurements between gray/white matter at 1.5 T (9.5-10.2), again increasing at higher fields (10.1-28.9 at 3 T and 10.9 at 4 T).

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。