Comparisons of automated blood pressures in a primary health care setting with self-measurements at the office and at home using the Omron i-C10 device

使用欧姆龙 i-C10 设备,比较基层医疗机构中自动血压测量值与诊室和家庭自测血压值。

阅读:1

Abstract

OBJECTIVE: We aimed to compare blood pressure (BP) levels recorded using the semiautomatic oscillometric Omron i-C10 BP device in patients with or without hypertension in three different settings: (a) when used by a doctor or a nurse at the office (OBP); (b) when used for self-measurement by the patient at the office (SMOBP); and (c) when used for 7 consecutive days at home (HBP). MATERIALS AND METHODS: A total of 247 individuals were invited to participate, but 78 of these individuals declined and a further seven were excluded, leaving a final cohort of 162 participants. RESULTS: The mean OBP was higher than HBP (difference 8.1±14/3.1±8.8 mmHg, P<0.0001) and so was SMOBP compared with HBP (difference 7.0±13/4.2±7.3 mmHg, P<0.0001). Sixteen participants (9.9%) had at least 10 mmHg higher systolic SMOBP than OBP and 28 (17%) participants had at least 10 mmHg lower systolic SMOBP than OBP. Participants who were current smokers had a larger mean difference between systolic OBP and SMOBP than nonsmokers (OBP-SMOBP in smokers: 6.6±9.4 mmHg, OBP-SMOBP in nonsmokers: 0.5±9.2 mmHg, P=0.011 between groups). CONCLUSION: Self-measurement of BP in the office does not preclude an increase in BP when levels in the individual patients are compared with HBP using the same equipment. Thus, SMOBP with a semiautomatic device does not lead to a reduction in the white-coat effect in the same manner as fully automatic devices.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。