Real-Time Navigation with Guide Template for Pedicle Screw Placement Using an Augmented Reality Head-Mounted Device: A Proof-of-Concept Study

利用增强现实头戴式设备进行椎弓根螺钉置入的实时导航及引导模板:概念验证研究

阅读:1

Abstract

OBJECTIVE: This study aims to explore the real-time navigation with guide template using an augmented reality head-mounted device (ARHMD) for pedicle screw placement. METHODS: The spatial coordinate relationships between augmented reality images and real objects were established through the custom-made guide template, and the registration and tracking were completed using an ARHMD. The feasibility and accuracy of this method were verified by pedicle screw placement in 2 lumbar models. According to the Gertzbein-Robbins grading scale, the accuracy of pedicle screw placement was assessed. The navigation errors were estimated by measuring the deviation values of entry point and trajectory angle. RESULTS: A total of 20 pedicle K-wires were placed into L1-L5 in 2 lumbar models, which were successfully completed, with an average time of 11.5 min per model and 69 s per screw. The overall K-wires placement accuracy was 100% (20 screws). The navigation error was 2.77 ± 0.82 mm for the deviation value of entry point, and 3.03° ± 0.94° for the deviation value of trajectory angle. CONCLUSIONS: The application of an ARHMD combined with guide template for pedicle screw placement is a promising navigation approach.