Pedicle screw piercer with warning device - A technique to increase accuracy of pedicle screw placement: A cadaveric study

带警示装置的椎弓根螺钉穿刺器——一种提高椎弓根螺钉置入精度的技术:一项尸体研究

阅读:1

Abstract

BACKGROUND: Pedicle screw fixation has achieved significant popularity amongst spinal surgeons for both single and multilevel spinal fusion. Suboptimal placements of pedicle screws may lead to neurological and vascular complications. There have been many advances in techniques available for navigating through the pedicle; however, these techniques are not without drawbacks. The purpose of this study was to investigate the efficacy and feasibility of the pedicle piercer with warning device. MATERIALS AND METHODS: Eight normal adult thoracolumbar specimens from cadavers consisting of 80 vertebras (T8-L5) were selected and randomly allocated into four groups. Each group contained 20 vertebra. Group 1 was tested for maximum pressure of the piercer within the vertebrae (F1). Group 2 was tested for maximum pressure of the warning piercer penetrating front cortex of the vertebral body (F2). Group 3 was tested for the maximum pressure of piercer penetrating vertebral body endplate (F3) and pedicle notch (F41, F42). Group 4 was tested for maximum pressure of the piercer penetrating the vertebral lateral cortex (F6), the medial and lateral cortex of pedicle (F51, F52). In the second experiment of this study, 4 normal adult specimens consisting of 40 vertebra and 80 pedicles were used for testing the alarm effects of pedicle piercer. The following indicators were adopted for the tests including true positive/negative, false positive/negative, sensitivity, specificity, availability, Youden index, and diagnostic efficiency. SPSS 16.0 was used for statistical analysis. RESULTS: There were statistically significant differences between F1, and F2, F3, F41, F42, F51, F52 respectively (P < 0.05). F1 = 8.970 ± 0.2698, F3 = 13.055 ± 0.6718. We found that the threshold value of piercer warning was from 9.6 to 12.3 Kgf. Sensitivity was 92.31%, specificity was 95.12%, usability was 87.45%, Youden index was 87.43% and diagnostic efficiency was 92.5% respectively. CONCLUSION: Warning piercer is a safe, simple, sensitive device for detecting pedicle breach during regular pedicle screw placement surgery.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。