Favorable results from a prospective evaluation of 200 patients with large hiatal hernias undergoing LINX magnetic sphincter augmentation

一项对 200 例接受 LINX 磁性括约肌增强术的大型食管裂孔疝患者进行的前瞻性评估取得了令人满意的结果。

阅读:1

Abstract

INTRODUCTION: Magnetic sphincter augmentation (MSA) of the lower esophageal sphincter restores the antireflux barrier in patients with hiatal hernias ≤3 cm. We performed a prospective study in patients undergoing MSA with the LINX device during repair of paraesophageal and hernias over 3 cm axial component. METHODS AND PROCEDURES: Multicenter, prospective study of consecutive patients treated with MSA at the time of repair of hiatal hernias >3 cm. RESULTS: 200 patients (110 female) were treated between March 2014 and February 2017 via laparoscopic hernia repair and MSA. Mean age was 59.5 years, mean BMI 29.4. 40% had esophagitis, 20% intestinal metaplasia, 72 of 77 tested had abnormal pH studies. Preoperative PPI use was reported by 87%. Eighteen patients had prior hiatal hernia/fundoplication. All had normal function. 78% of patients had axial hiatal hernia ≥5 cm or large paraesophageal component. Mean operative time was 81 min (38-193), EBL was 10 cc. Non-permanent mesh reinforcement of hiatal repair was performed in 83% of the patients. There were two readmissions for dehydration; 2 patients with pulmonary embolism, and 1 patient with cardiac ischemia. Nineteen patients required dilation. 156 pts were followed at a median of 8.6 months. GERD-HRQL scores improved from 26 preoperatively to 2 postoperatively. Complete PPI independence was achieved in 94% (147/156). Videoesophagram in 51 patients at median 11 months found 3 asymptomatic hernias <3 cm. One symptomatic patient underwent successful repair of the hernia without MSA manipulation. There have been no device explants, erosions, or migrations to date. CONCLUSIONS: This prospective study of 200 patients with >3 cm hernias undergoing MSA with hiatoplasty resulted in favorable outcomes with median of 9 months follow-up. Comparing this to published reports of MSA in patients with <3 cm hernias, the safety and clinical efficacy of MSA are independent of initial hernia size.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。