Determination of Mercury with a Miniature Sensor for Point-of-care Testing

利用微型传感器进行即时检测测定汞含量

阅读:1

Abstract

In developing countries, subsistence gold mining entails mixing metallic mercury with crushed sediments to extract gold. In this approach, the gold-mercury amalgam is heated to evaporate mercury and obtain gold. Thus, the highly volatile mercury can be absorbed through inhalation, resulting in adverse health effects. Urinalysis can be used to detect mercury, which is excreted in urine and feces, and correlate exposure with toxic effects. The current gold standard analytical methods are based on fluorescence or inductively coupled plasma mass spectrometry methods, but are expensive, time consuming, and are not easily accessible in countries where testing is needed. In this work, we report on a miniature electrochemical sensor that can rapidly detect mercury in urine at levels well below the US Biological Exposure Index (BEI) limit of 50 ppb (μg/L). The sensor is based on a thin-film gold electrode and anodic stripping voltammetry electroanalytical approach. The sensor successfully detected mercury at trace levels in urine, with a limit of detection of ~15 ppb Hg in the linear range of 20-80 ppb. With the low-cost disposable sensors and portable instrumentation, it is well suited for point-of-care applications.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。