Determination of manganese by cathodic stripping voltammetry on a microfabricated platinum thin-film electrode

采用阴极溶出伏安法在微加工铂薄膜电极上测定锰

阅读:3

Abstract

In this work, we report on the determination of trace manganese (Mn) using cathodic stripping voltammetry (CSV) using a microfabricated sensor with a Pt thin-film working electrode. While an essential trace metal for human health, prolonged exposure to Mn tends to gradually impair our neurological system. The potential sources of Mn exposure make it necessary to monitor the concentration in various sample matrices. Previous work by us and others suggested CSV as an effective method for measuring trace Mn. The analytical performance metrics were characterized and optimized, leading to a calculated limit of detection (LOD) of 16.3 nM (0.9 ppb) in pH 5.5, 0.2 M acetate buffer. Further, we successfully validated Mn determination in surface water with ~90% accuracy and >97% precision as compared with ICP-MS "gold standard" measurement. Ultimately, with stable, accurate and precise electrochemical performance, this Pt sensor permits rapid monitoring of Mn in environmental samples, and could potentially be used for point-of-use measurements if coupled with portable instrumentation.