Reducing applied force in colonoscopy using a novel soft robotic colonoscope: Head-to-head study

使用新型软体机器人结肠镜降低结肠镜检查中的施力:一项直接对比研究

阅读:1

Abstract

BACKGROUND AND STUDY AIMS: Current colonoscopies have a recognized limitation. Manual pushing and pulling required by operators provides gross movement in the bowel. Reported pain, therefore, is likely due to the manual force applied. Implementing novel steering techniques with a soft growing robotic system can potentially overcome challenges such as fine control, precise steering, and capability to expand treatment options for complex therapies. This study assessed a novel controlled-growing soft robot compared with a standard colonoscope in terms of force exhibited on a model bowel wall and its clinical implications. METHODS: A head-to-head study using a hybrid colon phantom of the left colon was undertaken. Both the novel soft-robot and standard colonoscope were passed through the phantom by endoscopists. Multiple passes were undertaken in the phantom with both colonoscopy methods with force values recorded at two points (rectum and sigmoid) with pressure sensors. RESULTS: Nine clinical endoscopists (4M:5F, 5 non-expert, 4 expert) were recruited. Average force with the novel robot was 0.25N (rectum) and 0.19N (sigmoid). Average force applied with standard colonoscopy was 2.82N (rectum) and 1.45N (sigmoid). CONCLUSIONS: This study demonstrated an improvement in force with the novel soft robot compared with a standard colonoscope. This suggests the possibility of more comfortable colonoscopy for patients. Currently time taken is longer with the novel robot, which is attributable to the learning curve and improves in subsequent passes. Further work will be undertaken in a complete colon model with aspirations to reach in-vivo experiments.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。