Low-dose pulsed vs standard pulsed fluoroscopy during ERCP to reduce radiation without change in image quality: Prospective randomized study

内镜逆行胰胆管造影术中低剂量脉冲透视与标准脉冲透视在降低辐射剂量且不影响图像质量方面的比较:前瞻性随机研究

阅读:1

Abstract

Background and study aims Endoscopic retrograde cholangiopancreatography (ERCP) poses the risk of radiation exposure (RE) to patients and staff and increases the risk of adverse biological effects such as cataracts, sterility, and cancer. Newer fluoroscopy equipment (C-Arm) provides options to limit radiation in the form of lower radiation dose and frame rate or time-limited "pulsed" settings. However, the impact of lower settings on image quality has not been assessed, and no standard protocol exists for fluoroscopy settings used during ERCP. Patients and methods This was a single-center, double-blind, prospective randomized study of consecutive adult patients undergoing standard-of-care ERCP at a tertiary academic medical center. Patients were randomized into two groups: 1) standard-dose pulsed and 2) low-dose pulsed. Pulsed mode (8 fps) was defined as x-ray exposure either in the manufacturer standard-dose or low-dose settings limited to 3 seconds each time the foot-operated switch was depressed. Results Seventy-eight patients undergoing ERCP were enrolled and randomized. No difference in age, gender, or body mass index was found between the two groups. No significant difference in image quality was found between standard-dose and low-dose fluoroscopy P = 0.925). The low-dose group was exposed to significantly less radiation when compared with standard-dose P < 0.05). Fluoroscopy time (minutes) was similar in both groups (2.0 vs 1.9), further suggesting that group assignment had no impact on image quality or procedure time. Conclusions Low-dose pulsed fluoroscopy is a reliable method that substantially reduces radiation without compromising image quality or affecting procedure or fluoroscopy times. This underscores the need for standardization in ERCP fluoroscopy settings to limit radiation exposure.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。