Usefulness of a laser-cut covered metal stent with a 7F delivery sheath in endoscopic ultrasound-guided biliary drainage without fistula dilation

激光切割覆膜金属支架(带7F输送鞘)在内镜超声引导下胆道引流(无需瘘管扩张)中的应用价值

阅读:1

Abstract

Background and study aims  Recently, the utility of endoscopic ultrasound-guided intervention without fistula dilation (EUS-IV WoD) has been reported to prevent adverse events. We clinically evaluated cases in which EUS-IV WoD was attempted using a novel self-expandable metallic stent (SEMS); this is a fully covered, laser-cut SEMS that has a tapered and stiff tip specifically designed for a 0.025-inch guidewire and a relatively thin, 7F delivery system. Patients and methods  We retrospectively evaluated cases wherein EUS-IV WoD was attempted using the novel SEMS between March and December 2021. Results  Treatment of 11 patients by EUS-IV WoD with the novel SEMS was attempted. The technical success rate for EUS-IV was 100 % and the clinical success rate was 100 %; the success rate for EUS-IV WoD was 72.8 %. Of these, the procedural success rate for EUS-IV WoD was 100 % in EUS-biliary drainage (BD) and 57.1 % in non-EUS-BD. Early adverse events were observed in 27.3 % of patients (3/11): mild abdominal pain in two patients and moderate bleeding in one patient. The abdominal pain cases were both cases of EUS-IV WoD failure and required fistula dilation. Conclusions  The novel stent may be useful for EUS-IV WoD, especially in EUS-BD.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。