A new device for endoscopic band ligation for colorectal diverticular bleeding

一种用于内镜下结直肠憩室出血套扎术的新型装置

阅读:1

Abstract

Background and study aims  Endoscopic band ligation (EBL) has been reported to be effective for colorectal diverticular bleeding. However, current EBL devices have some limitations, and we have developed a novel EBL device. Methods  This novel EBL device has a tapered hood attached to the tip of the endoscope and an outer cylinder that loads two EBL bands. Twelve EBL procedures were performed in a live porcine model alternately using a conventional EBL device (Group C) and the novel EBL device (Group N). Results  EBL was successful in all cases in both groups. There were no cases of perforation in the 10 days after EBL. After ligation, the mean major axis of the ridge was significantly larger in Group N than Group C (9.7 ± 1.4 mm vs 7.2 ± 1.4 mm, P  = 0.011). Pathological examination revealed disruption of the muscularis propria at four of the six ligation sites in Group C and at five of the six ligation sites in Group N. Conclusions  Using this novel EBL device, it was possible to perform multiple ligation procedures in succession with a good field of view. No perforation was observed, but disruption of the muscularis propria was observed at approximately three-quarters of the ligation sites pathologically.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。