Intraluminal gas escape from biopsy valves and endoscopic devices during endoscopy: caution advised during the COVID-19 era

内镜检查过程中活检阀和内镜装置内气体泄漏:新冠疫情期间需谨慎

阅读:1

Abstract

Background and study aims  The risk of aerosolization of body fluids during endoscopic procedures should be evaluated during the COVID-19 era, as this may contribute to serious disease transmission. Here, we aimed to investigate if use of endoscopic tools during flexible endoscopy may permit gas leakage from the scope or tools. Material and methods  Using a fresh 35-cm porcine rectal segment, a colonoscope tip, and manometer were placed intraluminally at opposite ends of the segment. The colonoscope handle, including the biopsy valve, was submerged in a water bath. Sequentially, various endoscopic devices (forceps, clips, snares, endoscopic submucosal dissection (ESD) knives) were inserted into the biopsy valve, simultaneously submerging the device handle in a water bath. The bowel was slowly inflated up to 74.7 mmHg (40 inH (2) O) and presence of gas leakage, leak pressure, and gas leakage volume were measured. Results  Gas leakage was observed from the biopsy valve upon insertion and removal of all endoscopic device tips with jaws, even at 0 mmHg (60/60 trials). The insertion angle of the tool affected extent of gas leakage. In addition, gas leakage was observed from the device handles (8 of 10 devices) with continuous gas leakage at low pressures, especially two snares at 0 mmHg, and an injectable ESD knife at 0.7 ± 0.8 mmHg). Conclusions  Gas leakage from the biopsy valve and device handles commonly occur during endoscopic procedures. We recommend protective measures be considered during use of any tools during endoscopy.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。