Intra-channel stent release technique for fluoroless endoscopic ultrasound-guided lumen-apposing metal stent placement: changing the paradigm

无透视内镜超声引导下腔内金属支架置入术的通道内支架释放技术:改变范式

阅读:1

Abstract

Background Recently, a novel lumen-apposing fully covered self-expanding metal stent (LA-FCSEMS) mounted on an electrocautery-enhanced delivery system has been developed to perform endoscopic ultrasound (EUS)-guided transluminal drainage. From early experience, however, release of the proximal flange of the stent has mostly been done using endoscopic view guidance to ensure proper positioning. Aim We describe a new technique that we have named the Intra-Channel Stent Release Technique (ICSRT) to perform stent placement under complete EUS control, without the use of either fluoroscopic or endoscopic views. Material and methods Data on all consecutive patients who underwent EUS-guided drainage using the new ICSRT between June 2014 and April 2016 were retrospectively retrieved from two institution databases. All EUS procedures were performed by experienced endoscopists with the patient under conscious or deep sedation. The total procedure and stent deployment time, and adverse events related to stent positioning with the ICSRT were evaluated. Results One hundred consecutive patients (51 women; mean age ± SD, 66 ± 15.2 years, range 34 - 95) underwent EUS-guided transluminal drainage with the Hot AXIOS™ device using the new ICSRT. The procedure was technically successful in all but one patient (1 %). The mean total procedural time was 21.9 minutes (range 7 - 50), while the mean time for stent placement was 3.2 minutes (range 1 - 15). No major adverse events occurred. Discussion The ICSRT has been used to deploy the newly developed lumen-apposing FCSEMS under complete EUS guidance without fluoroscopic and/or endoscopic assistance. The technique appears to be safe and highly effective and should be learned by all interventional endosonographers in order to be able to perform drainage in all clinical scenarios.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。