Improving Visualization of In-stent Lumen Using Prototype Photon-counting Detector Computed Tomography with High-resolution Plaque Kernel

利用高分辨率斑块核原型光子计数探测器计算机断层扫描技术改善支架内腔可视化

阅读:1

Abstract

The study aimed to compare the performance of photon-counting detector computed tomography (PCD CT) with high-resolution (HR)-plaque kernel with that of the energy-integrating detector CT (EID CT) in terms of the visualization of the lumen size and the in-stent stenotic portion at different coronary vessel angles. The lumen sizes in PCD CT and EID CT images were 2.13 and 1.80 mm at 0°, 2.20 and 1.77 mm at 45°, and 2.27 mm and 1.67 mm at 90°, respectively. The lumen sizes in PCD CT with HR-plaque kernel were wider than those in EID CT. The mean degree of the in-stent stenotic portion at 50% was 69.7% for PCD CT and 90.4% for EID CT. PCD CT images with HR-plaque kernel enable improved visualization of lumen size and accurate measurements of the in-stent stenotic portion compared to conventional EID CT images regardless of the stent direction.