Barlow's Transcatheter Edge-to-Edge Repair in Mitral Annular Calcification: Intracardiac Echocardiography-Guided

Barlow 经导管缘对缘修复二尖瓣环钙化:心内超声引导

阅读:1

Abstract

OBJECTIVE: The presence of Barlow's and mitral annular calcification poses significant challenges in mitral transcatheter edge-to-edge repair (M-TEER). Our case involved a patient with both pathologies who underwent successful 3-dimensional intracardiac echocardiography (ICE)-guided M-TEER technique with the PASCAL Precision system (Edwards Lifesciences). KEY STEPS: Our case was an 87-year-old woman with underlying percutaneous endoscopic gastrostomy feeding tube, failed transesophageal echocardiogram probe insertion, proceeded with 3-dimensional ICE-guided M-TEER. With fluoroscopic and ICE guidance, one PASCAL P10 was deployed at 12-6 o'clock lateral A2P2 position, a second PASCAL Ace was deployed at 11-5 o'clock at medial A2P2 position, and a third PASCAL Ace was deployed at 10-4 o'clock at A3P3 medial commissure position. Reduced mitral regurgitation occurred from severe to trivial-mild, without significantly increasing the mitral gradient. POTENTIAL PITFALLS: One of the tips in this case would be septostomy using the Armada balloon to facilitate manipulation of the ICE inside the left atrium. Careful manipulation of the delivery system under fluoroscopic guidance was also critical in multiple clips strategy, especially when starting the first clip in lateral position. A combination of P10 and Ace devices would likely be the choice in complex mitral anatomy cases. TAKE-HOME MESSAGE: Three-dimensional ICE-guided M-TEER with the use of P10 and Ace of PASCAL system in Barlow's with MAC is feasible with improvement in mitral regurgitation without elevating the mitral gradient.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。