Evaluation of Toxoplasma gondii immunoglobulin G (IgG) and IgM assays incorporating the newVidia analyzer system

采用新型Vidia分析仪系统对弓形虫免疫球蛋白G (IgG) 和IgM检测进行评估

阅读:1

Abstract

The new Vidia system is a fully automated system based on chemiluminescence and antigen bound to magnetic microparticles, which allows a fast measurement of Toxoplasma gondii-specific immunoglobulin G (IgG) and IgM levels. The analytical performances of the Vidia Toxo IgG and IgM assays were compared with those of the automated Vidas, AxSYM, and Liaison Toxo IgG and IgM assays. The comparative evaluation was performed utilizing 204 frozen sera belonging to 166 subjects and 201 fresh sera collected from 198 subjects. For the Vidia Toxo IgG system, the sensitivities were 100% in both the retrospective and prospective studies, and specificities were 98.39% in the retrospective study and 100% in the prospective study, respectively. The sensitivities of the other three Toxo IgG assays were 100%, and the specificities ranged from 96.77% to 100%. For the Vidia Toxo IgM assay, the sensitivity and specificity were 100% in both the retrospective and prospective studies. The overall sensitivities and specificities of the other three Toxo IgM assays ranged from 80% to 100% and from 99.44% to 100%, respectively. In our study, the Vidia system revealed excellent sensitivity (100% for both IgG and IgM assays) and good specificity (99.25% for IgG and 100% for IgM assays).

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。