Comparison of Ablation Performance between Octopus Multipurpose Electrode and Conventional Octopus Electrode

Octopus多功能电极与传统Octopus电极消融性能比较

阅读:1

Abstract

OBJECTIVE: To compare Octopus multipurpose (MP) electrodes, which are capable of saline instillation and direct tissue temperature measurement, and conventional electrodes for radiofrequency ablation (RFA) in porcine livers in vivo. MATERIALS AND METHODS: Sixteen pigs were used in this study. In the first experiment, RFA was performed in the liver for 6 minutes using Octopus MP electrodes (n = 15 ablation zones) and conventional electrodes (n = 12 ablation zones) to investigate the effect of saline instillation. The ablation energy, electrical impedance, and ablation volume of the two electrodes were compared. In the second experiment, RFA was performed near the gallbladder (GB) and colon using Octopus MP electrodes (n = 12 ablation zones for each) with direct tissue temperature monitoring and conventional electrodes (n = 11 ablation zones for each). RFA was discontinued when the temperature increased to > 60\xe2\x84\x83 in the Octopus MP electrode group, whereas RFA was performed for a total of 6 minutes in the conventional electrode group. Thermal injury was assessed and compared between the two groups by pathological examination. RESULTS: In the first experiment, the ablation volume and total energy delivered in the Octopus MP electrode group were significantly larger than those in the conventional electrode group (15.7 ± 4.26 cm(3) vs. 12.5 ± 2.14 cm(3), p = 0.027; 5.48 ± 0.49 Kcal vs. 5.04 ± 0.49 Kcal, p = 0.029). In the second experiment, thermal injury to the GB and colon was less frequently noted in the Octopus MP electrode group than that in the conventional electrode group (16.7% [2/12] vs. 90.9% [10/11] for GB and 8.3% [1/12] vs. 90.9% [10/11] for colon, p < 0.001 for all). The total energy delivered around the GB (2.65 ± 1.07 Kcal vs. 5.04 ± 0.66 Kcal) and colon (2.58 ± 0.57 Kcal vs. 5.17 ± 0.90 Kcal) were significantly lower in the Octopus MP electrode group than that in the conventional electrode group (p < 0.001 for all). CONCLUSION: RFA using the Octopus MP electrodes induced a larger ablation volume and resulted in less thermal injury to the adjacent organs compared with conventional electrodes.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。