Saline-enhanced hepatic radiofrequency ablation using a perfused-cooled electrode: comparison of dual probe bipolar mode with monopolar and single probe bipolar modes

采用灌注冷却电极的盐水增强肝脏射频消融术:双探针双极模式与单极和单探针双极模式的比较

阅读:1

Abstract

OBJECTIVE: To determine whether saline-enhanced dual probe bipolar radiofrequency ablation (RFA) using perfused-cooled electrodes shows better in-vitro efficiency than monopolar or single probe bipolar RFA in creating larger coagulation necrosis. MATERIALS AND METHODS: RF was applied to excised bovine livers in both bipolar and monopolar modes using a 200 W generator (CC-3; Radionics) and the perfused-cooled electrodes for 10 mins. After placing single or double perfused-cooled electrodes in the explanted liver, 30 ablation zones were created at three different regimens: group A; saline-enhanced monopolar RFA, group B; saline-enhanced single probe bipolar RFA, and group C; saline-enhanced dual probe bipolar RFA. During RFA, we measured the tissue temperature at 15 mm from the electrode. The dimensions of the ablation zones and changes in the impedance currents and liver temperature during RFA were then compared between the groups. RESULTS: The mean current values were higher for monopolar mode (group A) than for the bipolar modes (groups B and C): 1550+/-25 mA in group A, 764+/-189 mA in group B and 819+/-98 mA in group C (p < 0.05). The volume of RF-induced coagulation necrosis was greater in group C than in the other groups: 27.6+/-2.9 cm(3) in group A, 23.7+/-3.8 cm(3) in group B, and 34.2+/-5.1 cm(3) in group C (p < 0.05). However, there was no significant difference between the short-axis diameter of the coagulation necrosis in the three groups: 3.1+/-0.8 cm, 2.9+/-1.2 cm and 4.0+/-1.3 cm in groups A, B and C, respectively (p > 0.05). The temperature at 15 mm from the electrode was higher in group C than in the other groups: 70+/- 18 degrees C in group A, 59+/-23 degrees C in group B and 96+/-16 degrees C in group C (p > 0.05). CONCLUSION: Saline-enhanced bipolar RFA using dual perfused-cooled electrodes increases the dimension of the ablation zone more efficiently than monopolar RFA or single probe bipolar RFA.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。