Finite element study of the impact of pedicle screw density on the biomechanical response of a Lenke 1AN scoliotic curve

有限元研究椎弓根螺钉密度对Lenke 1AN型脊柱侧弯曲线生物力学响应的影响

阅读:1

Abstract

INTRODUCTION: Benefits of increasing screw density in posterior instrumentation used to treat a scoliotic deformity are demonstrated using a three-dimensional finite element model (FEM) of the thoracolumbosacral spine. METHODS: The FEM represents a Lenke 1AN scoliotic deformity with a 50° Cobb angle and 20° apical vertebral rotation. The curve is corrected with bilateral pedicle screw fixation and 75 separate randomized screw distributions. RESULTS: Total construct screw density, concave rod screw locations at T6, T10, T11 and T12, and convex rod screw locations at T7 and T12 each correlate strongly with reductions in postoperative Cobb angle (P < 0.05). Apical vertebral rotation is greatly impacted (reduced) by screws placed at the apical vertebra on both concave and convex rods (P < 0.05). Under pure moment loading, intersegmental micromotion is generally reduced when motion segment screw density is increased, with the exception being the upper instrumented joint. CONCLUSIONS: These results suggest that increasing the screw density of posterior constructs used to treat a Lenke 1AN scoliotic deformity may improve the de-rotation correction with better postural restoration, reducing the risk of future complications including pseudarthrosis.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。