Loading on Attune® fixed-bearing cruciate-substituting total knee implant in knee malalignment during activities of daily living: A finite element analysis

在日常生活活动中,膝关节畸形患者使用Attune®固定轴承式后交叉韧带替代全膝关节假体时的负荷情况:有限元分析

阅读:1

Abstract

PURPOSE: To compare contact stresses between Attune® and PFC Sigma® total knee arthroplasty (TKA) implants in the presence of knee malalignment. METHODS: Maximum contact stress after finite element analysis were compared during standing, walking, and stair climbing in 0°, 2.5° and 5°varus/valgus knee alignments. RESULTS: The percentage increase in contact stress was highest during walking with PFC Sigma® in 5° varus (238.5%), standing with Attune® in 5° valgus (127.2%), and standing with Attune® in 2.5° valgus (107.8%). CONCLUSION: The newer Attune® design may be associated with higher maximum contact stresses and increased risk of wear and implant failure.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。