Midterm results following uncemented, modular, fully porous coated stem used in revision total hip arthroplasty: Comparison of two stem systems

在翻修全髋关节置换术中使用非骨水泥、模块化、全多孔涂层柄的中期结果:两种柄系统的比较

阅读:1

Abstract

BACKGROUND: Aim of this study was to evaluate stem subsidence and survivorship of implant following implantation of Revitan (Zimmer) and MP (Waldemar Link) stems. METHODS: Retrospective case series with clinical and radiological follow-up of 1-10 years. RESULTS: 47 Revitan and 57 MP stems were analyzed. In 12 cases there was subsidence of the stem present. Two Revitan stems and one MP stem were revised as a result of subsidence. Failure rate was 4.3% for Revitan, 1.8% for MP stem and difference was not statistically significant. CONCLUSION: There is no significant difference in subsidence or survivorship between Revitan and MP stems.