Temperature changes caused by the difference in the distance between the ultrasound transducer and bone during 1 mhz and 3 mhz continuous ultrasound: a phantom study

MHz 和 3 MHz 连续超声作用下,超声换能器与骨骼之间距离差异引起的温度变化:一项体模研究

阅读:1

Abstract

[Purpose] This study aimed to use a thermograph to observe temperature changes caused by different distances between an ultrasound transducer and bone during 1 MHz and 3 MHz continuous ultrasound emission on a phantom. [Materials and Methods] We observed the distribution of temperature elevations on a phantom consisting of pig ribs and tissue-mimicking material. One megahertz and 3 MHz ultrasound were delivered at 2.0 W/cm(2) for 5 minutes. To record the temperature changes on the phantom, we took a screenshot of the thermograph with a digital camera every 20 seconds. [Results] With 1 MHz ultrasound at the distances of 2 and 3 cm, the temperature elevation near the bone was higher than that near the transducer. However, with 3 MHz ultrasound, the temperature elevation was higher near the transducer rather than near the bone. At this point, we consider that there is a possibility of heat injury to internal organs in spite of there being no elevation of skin temperature. [Conclusion] When performing ultrasonic therapy, not only should the frequency be taken into consideration, but also the influence of the absorption coefficient and the reflection of the tissue. We visually confirmed the thermal ultrasound effect by thermography. Special attention to the temperature elevation of the internal organs is necessary to avoid injuries.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。