Concurrent Validity and Test-retest Reliability of the OPTOGait Photoelectric Cell System for the Assessment of Spatio-temporal Parameters of the Gait of Young Adults

OPTOGait光电细胞系统在评估青年人步态时空参数方面的同时效度和重测信度

阅读:1

Abstract

[Purpose] The purpose of this study was to investigate the concurrent validity and test-retest reliability of the recently introduced OPTOGait Photoelectric Cell System for the assessment of spatio-temporal parameters of gait. [Subjects] Twenty healthy young adults (mean age = 27.35, SD = 7.4) were asked to walk 3 times on walkway at a comfortable speed. [Methods] Concurrent validity was assessed by comparing data obtained using the OPTOGait and GAITRite systems, and reliability was assessed by comparing data from the first and third OPTOGait sessions. [Results] Concurrent validity, as identified by intra-class correlation coefficients (ICC (2, 1) = 0.929-0.998), coefficients of variation (CVME = 0.32-11.30%), and 95% limits of agreement, showed high levels of correlation. In addition, the test-retest reliability of the OPTOGait Photoelectric Cell System was demonstrated as showing a high level of correlation with all spatio-temporal parameters by intra-class correlation coefficients (ICC (3, 1) = 0.785-0.952), coefficients of variation (CVME = 1.66-4.06%), 95% limits of agreement, standard error of measurement (SEM = 2.17-5.96%), and minimum detectable change (MDC95% = 6.01-16.52%). [Conclusion] The OPTOGait Photoelectric Cell System has strong concurrent validity along with relative and absolute test-retest reliabilities. This portable system with easy-to-use features can be used for clinical assessments or research purposes as an objective means of assessing gait.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。