Accuracy of facial skeletal surfaces segmented from CT and CBCT radiographs

基于CT和CBCT影像分割的面部骨骼表面的准确性

阅读:1

Abstract

The accuracy of three-dimensional (3D) facial skeletal surface models derived from radiographic volumes has not been extensively investigated yet. For this, ten human dry skulls were scanned with two Cone Beam Computed Tomography (CBCT) units, a CT unit, and a highly accurate optical surface scanner that provided the true reference models. Water-filled head shells were used for soft tissue simulation during radiographic imaging. The 3D surface models that were repeatedly segmented from the radiographic volumes through a single-threshold approach were used for reproducibility testing. Additionally, they were compared to the true reference model for trueness measurement. Comparisons were performed through 3D surface approximation techniques, using an iterative closest point algorithm. Differences between surface models were assessed through the calculation of mean absolute distances (MAD) between corresponding surfaces and through visual inspection of facial surface colour-coded distance maps. There was very high reproducibility (approximately 0.07 mm) and trueness (0.12 mm on average, with deviations extending locally to 0.5 mm), and no difference between radiographic scanners or settings. The present findings establish the validity of lower radiation CBCT imaging protocols at a similar level to the conventional CT images, when 3D surface models are required for the assessment of facial morphology.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。