Influence of CBCT-based volumetric distortion and beam hardening artefacts on the assessment of root canal filling quality in isthmus-containing molars

CBCT体积畸变和束硬化伪影对含峡部磨牙根管充填质量评估的影响

阅读:1

Abstract

OBJECTIVES: To evaluate the influence of artefacts in cone beam CT (CBCT) images of filled root canals in isthmus-containing molars. METHODS: 10 teeth presenting canals with an isthmus were instrumented and filled with a thermoplasticised obturation technique. The teeth were scanned using a micro-CT device and two CBCT devices: 3D Accuitomo 170 (ACC) and NewTom VGi evo (NT), with different acquisition protocols: larger and smaller voxel size. Three examiners assessed the CBCT images for: (1) detection of filling voids; (2) assessment of under- or overestimation of the filling material and (3) resemblance of CBCT images to the reference standard. Analyses of Task 1 yielded accuracy, sensitivity and specificity for detection of filling voids. For tasks 2 and 3, statistical analysis was performed using Wilcoxon test. The level of significance was set at p < .05. RESULTS: For Task 1, ACC showed higher sensitivity, whereas NT presented higher specificity. No significant difference was found between the protocols in ACC, however, for NT, differences between protocols were significant for all diagnostic values. In Task 2, visualisation of the filling was overestimated for NT, while for ACC, underestimation was observed. For Task 3, images with smaller voxel size were more similar to the reference image, for both CBCT devices. CONCLUSIONS: Different artefacts compromise the detection of filling voids on CBCT images of canals in mandibular molars with isthmus. ACC and NT present rather similar diagnostic accuracy, even though artefact expression remains device-specific.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。