Re-exposure in cone beam CT of the dentomaxillofacial region: a retrospective study

口腔颌面部锥形束CT重复曝光:一项回顾性研究

阅读:1

Abstract

METHODS: In a retrospective cohort study CBCT images of 4986 patients from the patient database from the Department of Oral Radiology, University Medical Center of the Johannes Gutenberg University, Mainz, Germany were included and the rate of re-exposures was counted. Patients were stratified into those who received a scan with the small field-of-view CBCT or the large field-of-view CBCT. The effect of patient-related parameters as age and gender was implicated. As a further device-specific parameter, the statistical analysis included whether the selection of the field of view due to the device type had a significant influence on the occurrence of re-exposures. Furthermore, the rescans were analyzed with regard to their causes. RESULTS: In total, CBCT images of 82 (1.6%) patients had to be repeated. Looking at the two different devices, in 42 (1.3%) patients that received a scan with the large field-of-view CBCT and in 40 (2.3%) patients that received a scan with the small field-of-view CBCT respectively needed a retake. There was no statistically significant correlation between age and gender to retakes. For the small field-of-view-size significantly more retakes were observed than for the large one. With 46% motion artifacts were the most frequent causes for a re-exposure of the patient. CONCLUSIONS: Gender and age did not have an impact on the occurrence of re-exposures. Patients who received a scan with the small field-of-view CBCT were significantly more often rescanned than those with the large field-of-view CBCT.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。