The effect of computed tomographic scanner parameters and 3-dimensional volume rendering techniques on the accuracy of linear, angular, and volumetric measurements of the mandible

计算机断层扫描仪参数和三维容积渲染技术对下颌骨线性、角度和体积测量精度的影响

阅读:1

Abstract

OBJECTIVES: This study investigates the effect of scanning parameters on the accuracy of measurements from three-dimensional (3D), multi-detector computed tomography (MDCT) mandible renderings. A broader range of acceptable parameters can increase the availability of computed tomographic (CT) studies for retrospective analysis. STUDY DESIGN: Three human mandibles and a phantom object were scanned using 18 combinations of slice thickness, field of view (FOV), and reconstruction algorithm and 3 different threshold-based segmentations. Measurements of 3D computed tomography (3DCT) models and specimens were compared. RESULTS: Linear and angular measurements were accurate, irrespective of scanner parameters or rendering technique. Volume measurements were accurate with a slice thickness of 1.25 mm, but not 2.5 mm. Surface area measurements were consistently inflated. CONCLUSIONS: Linear, angular, and volumetric measurements of mandible 3D MDCT models can be confidently obtained from a range of parameters and rendering techniques. Slice thickness is the primary factor affecting volume measurements. These findings should also apply to 3D rendering using cone-beam CT (CBCT).