Burst Pressures in Eyes with Clear Corneal Incisions Treated with ReSure Glue

使用 ReSure 胶水治疗透明角膜切口后眼内爆裂压

阅读:2

Abstract

PURPOSE: This study aims to measure burst pressures in 3 mm clear corneal incisions sealed with ReSure, a biodegradable hydrogel sealant, and to compare it to traditional 10-0 nylon sutures and unsealed controls. DESIGN: An ex vivo animal study. METHODS: 3 mm clear corneal incisions were performed in rabbit eyes (ex vivo). The burst pressure was determined, and then, the incisions were sealed with either ReSure glue or a single 10-0 nylon suture. Burst pressure measurements were repeated. RESULTS: Fourteen eyes were included. The median burst pressure in the suture-control group (7 eyes) prior to suture application was 7 mmHg (range: 0-45); the median burst pressure in the 7 glue-controls was 36 mmHg (range: 5-61, p = 0.08 for the comparison of the two control groups). The median burst pressure in the glue group was 93 mmHg (range: 39-129, p = 0.043 when compared to glue-control). The median burst pressure in the suture group was 158 mmHg (range: 70-180, p = 0.018 when compared to suture-control). There was no statistically significant difference in burst pressure values between the glue and suture groups (p = 0.08). CONCLUSION: In this study, ReSure glue applied to 3 mm clear corneal incisions provided sufficient resistance to elevated intraocular pressure when compared to controls. The results of this study suggest that ReSure glue may be comparable to a single 10-0 nylon suture in resisting fluid egress during the early postoperative period.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。