In Vitro Aberrometric Assessment of a Multifocal Intraocular Lens and Two Extended Depth of Focus IOLs

体外像差分析多焦点人工晶状体和两种景深扩展型人工晶状体

阅读:2

Abstract

PURPOSE: To analyze the "in vitro" aberrometric pattern of a refractive IOL and two extended depth of focus IOLs. METHODS: A special optical bench with a Shack-Hartmann wavefront sensor (SH) was designed for the measurement. Three presbyopia correction IOLs were analyzed: Mini WELL (MW), TECNIS Symfony ZXR00 (SYM), and Lentis Mplus X LS-313 MF30 (MP). Three different pupil sizes were used for the comparison: 3, 4, and 4.7 mm. RESULTS: MW generated negative primary and positive secondary spherical aberrations (SA) for the apertures of 3 mm (-0.13 and +0.12 μm), 4 mm (-0.12 and +0.08 μm), and 4.7 mm (-0.11 and +0.08 μm), while the SYM only generated negative primary SA for 4 and 4.7 mm apertures (-0.12 μm and -0.20 μm, resp.). The MP induced coma and trefoil for all pupils and showed significant HOAs for apertures of 4 and 4.7 mm. CONCLUSIONS: In an optical bench, the MW induces negative primary and positive secondary SA for all pupils. The SYM aberrations seem to be pupil dependent; it does not produce negative primary SA for 3 mm but increases for higher pupils. Meanwhile, the HOAs for the MW and SYM were not significant. The MP showed in all cases the highest HOAs.