Autofluorescence Images with Carl Zeiss versus Topcon Eye Fundus Camera: A Comparative Study

使用卡尔蔡司和拓普康眼底相机拍摄的自发荧光图像:一项比较研究

阅读:1

Abstract

Purpose. To compare the autofluorescence images of the Zeiss versus Topcon eye fundus cameras and design an objective way to quantify it. Procedures. The IMAGEJ software was used to determine the gray level corresponding to the darkest veins and the peripapillary ring (thresholds), the level of white of the brightest perifoveal area, their difference (contrast level), and the suprathreshold area for each photograph. Results. Carl Zeiss has higher contrast values than Topcon. The Topcon contrast presented a crest with further decline as the suprathreshold area continued to increase. On the contrary, the Zeiss profile did not decline in contrast. Conclusions and Message. The Carl Zeiss camera showed superior contrast ability over the Topcon when performing autofluorescence imaging. We set objective parameters to compare fundus cameras FAF images. These parameters could be the base to objectively measure and determine changes and realize followup to areas of hyper- or hypofluorescence.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。