Flow through a mechanical distraction enterogenesis device: a pilot test

机械性肠扩张装置的流动情况:一项初步测试

阅读:1

Abstract

BACKGROUND: We tested the coupling portion of a prototype intraluminal distraction enterogenesis device to allow flow-through of simulated enteric contents (SEC) in both pig and human jejunum. MATERIALS AND METHODS: SEC was made using 80% corn syrup. Ten-cm pig and human intestinal segments had a spoke-shaped 2.2 cm coupling adaptor sutured in place, intraluminally. The adaptor had a flow-through area of 33.6 mm(2). SEC was pumped into the proximal part of the intestinal segment at 0.083 mL/s. The times to first passage of SEC through the coupler (first drop), 10 mL, and 20 mL of SEC eluted from the distal end were recorded. RESULTS: Mean time to first drop elution was 155 ± 38 s with pig, and 149 ± 22 s with human bowel (P = 0.8). This corresponded to a hydrostatic pressure of 37.5 mmHg before the initial drop passed through. Mean flow rates were 0.094 mL/s in pig bowel and 0.084 mL/s in human bowel (P = 0.09). To account for occlusion from luminal debris, a 75% occlusion of coupler holes was studied in the smaller pig bowel to investigate if reductions in flow-through area could be tolerated. Mean time to first drop increased slightly to 171 ± 15 s, but the elution rate stayed the same (P = 0.5). CONCLUSIONS: After a physiologic level of initial pressure buildup allowing the first drop of SEC to pass the coupling adaptor, our prototype intestinal coupling adaptor did not obstruct flow-through of SEC, even after a 75% decrease in flow-through area. This type of attachment represents a viable approach to placing a device in-continuity without obstructing flow of enteric contents.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。